Procedure
Pacemaker
search
Pacemaker
, Twiddler's Syndrome, Pacemaker Interrogation
See Also
Implantable Defibrillator
(AICD)
Cardiac Resynchronization Implantable Device
Left Ventricular Assist Device
(
LVAD
)
Cardiac Pacemaker Infection
Definition
Thoracic-implanted, self-contained (with battery and circuitry)
Cardiac sensing and pacing device indicated for
Symptomatic Bradycardia
or
Syncope
Indications
Permanent Pacemaker -
Gene
ral
Sinus Node Dysfunction
with
Symptomatic Bradycardia
or
Syncope
Advanced second or third degree
AV Block
(see below)
Chronic bifascicular block
Acute
Myocardial Infarction
requiring ventricular pacing beyond acute phase
Hypersensitive carotid sinus syndrome and
Neurocardiogenic Syncope
Cardiac Transplantation
requiring post-operative pacing
Symptomatic recurring SVT
Congenital long
QT Interval
Cardiac Resynchronization Therapy
with biventricular pacing
Symptomatic
Heart Block
with
Cardiomyopathy
or
Congenital Heart Disease
Indications
Permanent Pacemaker -
AV Block
Class I Indications (helpful)
Sick Sinus Syndrome
Symptomatic Bradycardia
with frequent sinus pauses (>3 seconds) associated with symptoms
Symptomatic chronotropic incompetence (inadequate
Heart Rate
response to
Exercise
or activity)
Second Degree
AV Block
with
Symptomatic Bradycardia
Third Degree
AV Block
with one associated condition
Symptomatic Bradycardia
Documented
Asystole
(3 seconds or greater)
Catheter ablation of the AV junction
Neuromuscular disorder with
AV Block
Myotonic muscular disorder
Kearns-Sayre Syndrome
Erb's Dystrophy (limb-girdle)
Peroneal muscular atrophy
Class IIa Indications (probably helpful)
Sick Sinus Syndrome
Symptomatic Bradycardia
with
Heart Rate
<40 bpm but symptom association with
Bradycardia
unclear
Idiopathic
Syncope
and
Sinoatrial Node
dysfunction identified on electrophysiologic studies
Asymptomatic third degree
AV Block
Asymptomatic Type II second degree
AV Block
Asymptomatic Type I
AV Block
at His level
First degree
AV Block
and Pacemaker syndrome symptoms
Class IIb Indications (Possibly helpful)
Sick Sinus Syndrome
Chronic awake rate <40 bpm in minimally symptomatic patients
Marked First Degree
AV Block
(>0.3 seconds) with CHF
Hypersensitive carotid sinus syndrome with recurrent
Syncope
ACC/AHA recommends if associated with ventricular
Asystole
for 3 or more seconds
Cochrane review does not find sufficient evidence supporting Pacemakers for
Carotid Sinus Syncope
Romme (2011) Cochrane Database Syst Rev (10): CD004194 [PubMed]
Class III Indications (Not helpful, possibly harmful)
Asymptomatic
Bradycardia
due to medication
Asymptomatic First Degree
AV Block
Asymptomatic Type I
AV Block
limited to supra-His
Transient
AV Block
secondary to resolving condition
Drug toxicity
Lyme Disease
Components
Pulse
generator (Electronic device)
Battery (e.g. 5-10 year
Lithium
ion)
Pacing leads
Description
Pacemaker Codes
Background
Most Pacemakers in the U.S. are DDD (dual lead, dual sensed, dual response of trigger and inhibit)
Some Pacemakers placed outside the U.S. are VVI (ventricle paced, ventricle sensed, ventricle inhibited)
Position 1 (chamber paced)
V - Ventricle
A - Atrium
D - Dual (A and V)
O - None
Position 2 (chamber sensed)
V - Ventricle
A - Atrium
D - Dual (A and V)
O - None
Position 3 (response to sensing)
V - Triggered
Intrinsic
P Wave
with no QRS triggers the Pacemaker
I - Inhibited
Intrinsic
P Wave
with a QRS inhibits the Pacemaker
D - Dual (T and I)
O - None
Position 4 (programmable functions and rate modulation)
P - Programmable rate and output
M - Muti-programmability of rate, output, sensitivity
C - Communicating via telemetry
R - Rate modulation
O - None
Position 5 (anti-tachyarrhythmia)
P - Pacing (anti-tachyarrhythmia)
S -
Shock
D - Dual (P and S)
O - None
Types
Pacemaker modes
Synchronous mode
Synchronous with heart beat
Asynchronous mode (fixed mode)
Asynchronous in regards to heart beat (no response to sensing)
Pacemakers revert to asynchronous mode when exposed to a magnet
Types
Pacemaker Selection for
Sinus Node Dysfunction
No signs or future risks for impaired AV conduction
Rate response: Rate-responsive atrial pacer (AAIR)
No rate response: Atrial Pacemaker (AAI)
Impaired AV Conduction and no AV synchrony needed
Rate response: Rate-responsive dual chamber (DDDR)
No rate response: Ventricular Pacemaker (DDD)
Impaired AV Conduction and AV synchrony needed
Tachyarrhythmia (e.g.
PSVT
)
Rate response: Rate-responsive dual chamber (DDDR)
No rate response: Dual chamber Pacemaker (DDD)
No Tachyarrhythmia
Rate response: Rate-responsive dual and mode switch
No rate response: Dual chamber with mode switching
Precautions
Magnetic field exposure
Pacemakers typically switch to asynchronous pacing at a set rate on exposure to magnetic field
Cell phones should be held on the opposite side of body, away from Pacemaker
Had been theoretical risk only in the past
More powerful magnets are as of 2022 installed in phones for wireless charging (e.g. IPhone 12 MagSafe)
Case reports of magnet triggered
Syncope
and other pacer adverse events are now more common
Nadeem (2021) J Am Heart Assoc
https://www.ahajournals.org/doi/10.1161/JAHA.121.020818
Other wearable and personal use devices (e.g.
Electronic Cigarette
s) may interfere with Pacemaker function
Shea (2020) HeartRhythm Case Rep 6(3): 121-3 [PubMed]
Asher (2020) HeartRhythm Case Rep 7(3): 167-9 [PubMed]
Magnets may be helpful in some emergency settings (e.g. applied to AICD that is delivering inappropriate shocks)
MRI scans have been historically contraindicated
Devices manufactured after 2000 are considered safe for non-thoracic MRI (after first 6 weeks)
Avoid MRI with 6 weeks of placement due to device dislodgement
Set Pacemaker to asynchronous mode (reset after scan), and turn AICD devices off
Pacemaker may distort images
Typically performed at tertiary centers with emergency backup and temporary Pacemaker reprogramming
Russo (2017) N Engl J Med 376(8): 755-64 +PMID:28225684 [PubMed]
Nazarian (2013) Circ Arrhythm Electrophysiol 6(2): 419-28 [PubMed]
Battery life remaining in Pacemaker
Pacers are interrogated to determine remaining battery life
When battery life drops below ERI (Elective replacement indicator),
Heart Rate
will be be fixed at a manufacturer-set rate
When battery life drops below EOL (End of life),
Heart Rate
will be fixed at a different fixed manufacturer-set rate
External
Defibrillation
, cardioversion and external pacing
Electrical shock may theoretically damage Pacemaker
Emergency
Defibrillation
may be performed without regard to Pacemaker
Avoid applying elective cardioversion pads directly over Pacemaker
Try to apply pads at least 10-15 cm away from the device
Imaging
Chest XRay
Indications
May identify misplaced or damaged lead
Identifies the Pacemaker type via a radiodense device stamp (requires a high penetration
Chest XRay
)
Interpretation - Normal Pacemaker wire configuration
Most newer Pacemakers are dual lead (but some may have only a single lead to either atrium or ventricle)
Pacemaker tip positioning - PA View (or AP View)
Pacemaker leads should coss the midline (from the left chest wall implantation site to the right heart)
Atrial leads should be directed slightly upwards
Ventricular leads should be directed slightly downwards
Pacemaker tip positioning - Lateral View (or Lateral View)
Pacemaker leads should be directed towards the
Sternum
(anteriorly)
As with PA view, atrial leads are directed upward and ventricular leads downward
Diagnostics
Electrocardiogram
Pacer spike
Long, very narrow signal preceding a complex
ST and
T Wave
s
Should always be discordant or opposite to the major
QRS Complex
(similar to a
Left Bundle Branch Block
)
Discordance (or
QRS Complex
in the opposite direction as the
ST Segment
and
T Wave
) is normal
Concordance (QRS and ST/
T Wave
in same direction) may suggest acute
Myocardial Infarction
Sgarbossa Criteria
may distinguish Pacemaker altered ST-T morphology versus an acute
Myocardial Infarction
Applies to
Left Bundle Branch Block
induced with a right ventricular Pacemaker lead
No guidelines directing
Sgarbossa
use in Pacemaker patients
However original
Sgarbossa
study included a subset of Pacemaker patients
QRS Complex
Typically
Left Bundle Branch Block
(Pacemaker leads are typically placed in the right ventricle)
Diagnostics
Magnet placed over Pacemaker
Indicated if Pacemaker malfunction is suspected
Perform an
Electrocardiogram
before and after applying magnet
Before magnet application (synchronous mode)
Pacing is typically rate responsive via Pacemaker sensing functionality
After magnet application (asynchronous mode, fixed mode)
Turns off Pacemaker sensing functionality
Pacemaker will now pace at an intrinsic rate (80-100 depending on manufacturer)
QRS Complex
es present after magnet application suggests a failure of Pacemaker sensing
No
QRS Complex
es after magnet applications may suggest a failure of Pacemaker capture
Diagnostics
Pacemaker Interrogation
All Pacemakers can perform
Electrocardiogram
s (may be turned off to conserve battery life)
Electrocardiogram
functionality is enabled during interrogation
Pacemaker programming head is placed directly over Pacemaker and push interrogate button
Identify the patient's underlying rhythm
Are the pacer spikes atrial or ventricular?
Is the rhythm Pacemaker dependent?
If uncertain, Pacemaker rate can be slowed to see if patient's rate also drops
Can the Pacemaker sense the heart rhythm?
Can the Pacemaker pace the heart?
Are there other rhythm problems?
Some Pacemakers can store abnormal rhythm events (e.g. VT runs) with a date-time stamp
Is the Pacemaker programmed correctly?
ERI (Elective replacement indicator)
Battery life remaining before replacement is needed
Management
Troubleshooting a malfunctioning Pacemaker
See
Cardiac Pacemaker Infection
Evaluation
See
Chest XRay
(above) for Pacemaker lead positioning and device stamp
See Magnet Application (above)
See
Electrocardiogram
(above) for Pacemaker-related findings
See Pacemaker Interrogation (above)
Pacemaker device malfunction categories
Failure to capture (despite pacer spike)
Pacemaker spike present without ensuing
QRS Complex
May indicate lead
Fracture
or dislodgement, or malfunction central to the device
Lead displacement is the most common cause and typically occurs in first month
Failure to output (without pacer spike, includes oversensing, exit block)
Pacer spikes are absent
May indicate lead
Fracture
or dislodgement, battery depletion
May indicate device oversensing
Pacemaker inhibited by a
Hyperacute T Wave
, muscle
Fasciculation
, cell phone within 10 cm
Misinterpreted as an intrinsic beat
Applying a magnet over the Pacemaker is diagnostic and therapeutic
May indicate exit block
Altered interface between Pacemaker lead and endocardium
Causes include MI, fibrosis,
Hyperkalemia
,
Antiarrhythmic
s
Failure to sense (undersensing)
Pacemaker fails to sense intrinsic cardiac activity and paces as if intrinsic activity is absent
Pacer spike appear regardless of intrinsic rhythm
May indicate lead dislodgement, altered position, acute right ventricular infarction
Risk of R on T phenomenon
Response depends on Pacemaker mode and programmed sensing parameters
Placing a magnet over the Pacemaker will disable the aberrant functionality until it may be repaired
Specific Pacemaker problems
Dislodged/displaced or
Fracture
d Pacemaker lead
Pacemaker lead displacement is most common
Pacemaker leads are most prone to
Fracture
at the lead insertion site and under the clavicle or first rib
Pacemaker syndrome (up to 20% of cases)
Loss of atrial capture with only ventricle paced (atria contracts against a closed AV valve)
Presents with retrograde pulsations into the neck
Atrial distention results in diuresis and
Hypotension
Loss of
Preload
may result in secondary
Heart Failure
Symptoms include weakness,
Fatigue
,
Dizziness
,
Orthostasis
,
Dyspnea
and
Chest Pain
Runaway Pacemaker syndrome
A damaged Pacemaker theoretically could sporadically increase paced rates at well above 100
Pacemaker mediated
Tachycardia
(endless loop
Tachycardia
) is a reentrant variant
Rare, but life threatening
Typically due to battery failure or Pacemaker damage
Apply magnet to deactivate sensing and terminate reentrant
Dysrhythmia
Typically requires emergent removal of the device
Inadequate pacer energy
Twiddler's Syndrome
Pacemaker patients may manipulate the Pacemaker site (subconsciously)
Results in lead retraction and displacement
Presentations
Pacemaker failure (e.g.
Syncope
)
Symptoms specific to new lead location (e.g.
Hiccup
s, diaphragmatic spasm, arm twitching)
Risk factors
Elderly or obese (loose subcutaneous Pacemaker pocket)
Dementia
,
Obsessive Compulsive Disorder
or
Developmental Delay
Diagnosis
Electrocardiogram
(evaluate for appropriate sensing)
See
Chest XRay
below regarding lead positioning (also observe for
Fracture
d leads)
References
Jhun and Shoenberger in Herbert (2015) EM:Rap 15(7): 17
Data to have available when communicating with Cardiology about a patient with possible Pacemaker-related problem
Pacemaker information
Most patients will have a card with Pacemaker information (device type, interrogation phone numbers)
Pacemaker type can also be read from a high penetration
Chest XRay
Is the device a Pacemaker or
Defibrillator
?
How many wires to heart are present?
When and where was the device implanted?
Precautions
Magnetic Resonance Imaging
(MRI)
Most Pacemakers can not undergo MRI
Risk of
Arrhythmia
or capture loss
Certain tertiary centers perform MRI on non-MRI conditional devices with electrophysiologist present
(2017) Am J Neuroradiol 38:2222 [PubMed]
Abandoned leads can not typically undergo MRI
Lead super-heating and local myocardial wall ablation
MR Conditional Devices and leads may undergo MRI under certain conditions
Most manufacturers as of 2018 produce MR conditional devices
Obtain
Chest XRay
in advance of MRI
Requires Pacemaker temporary reprogramming (off or asynchronous mode with higher rates and outputs)
Expect scatter on imaging from the device (for targets near the device)
References
Abresch (2018)
Cardiac Arrhythmia
Conference, UMN, Minneapolis
Complications
Acute complications associated with placement
Pneumothorax
or
Hemothorax
(complicates 1-3% of Pacemaker placements)
Myocardial perforation (with associated
Cardiac Tamponade
risk)
Brachial nerve injury
Pacemaker Infection
(1-7% risk)
Acute lead dislodgment (2-4% risk)
Complications
Chronic complications
Atrial Fibrillation
Increased risk with ventricular pacing (e.g. RV pacing)
References
Jones and Orman in Majoewsky (2012) EM:Rap 12(5): 4-6
Mallemat, Swaminathan and Egan in Herbert (2014) EM:Rap 14(10): 5-7
Vanlandingham (2015) Crit Dec Emerg Med 29(10): 2-14
Bernstein (1987) Pacing Clin Electrophysiol 10:794-9 [PubMed]
Denay (2014) Am Fam Physician 89(4): 279-82 [PubMed]
Epstein (2008) J Am Coll Cardiol 51(21):e1-62 [PubMed]
Gregoratos (1998) J Am Coll Cardiol 31:1175-209 [PubMed]
Gregoratos (2005) Am Fam Physician 71:1563-70 [PubMed]
Type your search phrase here