Procedure
Endoscope Cleaning
search
Endoscope Cleaning
Immediately after procedure
Immediately immerse scope tip in water after procedure
Flush suction channel to prevent clogging with debris
Brush suction channel and flush suction channel again
Place scope tubing in mild detergent solution
Scope cleaning
Wash scope external sheath with sponge
Remove and clean biopsy, suction, air and water valves
Clean biopsy channel with mild detergent solution
Rinse scope with tap water and flush all channels
Clean biopsy forceps
Scope Disinfection
Suction 50 ml Glutaraldehyde through biopsy channel
Fill water insufflation bottle >50 ml glutaraldehyde
Flush glutaraldehyde through air/water channel
Lay scope tubing in glutaraldehyde for 10-20 minutes
Lay biopsy forceps in glutaraldehyde for 10-20 minutes
Do not immerse control head
Scope Rinsing
Rinse scope in tap water
Flush biopsy channel for 60 seconds with warm tap water
Rinse scope tubing with tap water
Suction air through biopsy channel for 60 seconds
Suction air through air/water channel for 60 seconds
Dry valves with cotton-tipped swabs
Scope Drying
Hang scope to dry
Leave water bottle open to air dry